Safety and Efficacy Advances in Surgical Tourniquets

Posted on November 22, 2013. Filed under: Safety and Efficacy Advances | Tags: , , , , , , , , , , , , |

Jim McEwen PhD

 

New AORN Recommended Practices for Surgical Tourniquets

All surgical tourniquet users should be aware that on June 15, 2013 the AORN (Association of periOperative Registered Nurses) published a major update of its recommended practices on pneumatic tourniquet used in surgery.   These guidelines are widely used and followed, especially in accreditation of surgical facilities.  Many changes in recommended practices have been made, and new topics have been added, as outlined below and as given at http://www.tourniquets.org .   The RP is available for purchase at http://aornstandards.org/.

 

Comparative Effectiveness  

Outpatient surgical staff and facilities are paying increased attention to ‘comparative effectiveness’.   Comparative effectiveness research (CER) is a method of comparing different devices and treatments to determine which is the most effective.  The growth of CER highlights the value of research and published evidence quantifying the benefits of medical devices, and a growing body of evidence supports recent advances seen in surgical tourniquet instruments, cuffs and accessories.  Updated information on recent publications and evidence can be found at www.tourniquets.org and in the 2013 AORN Recommended Practices.

Two disturbing developments related to comparative effectiveness have occurred recently.  First, counterfeit tourniquet products have been seen. These counterfeit products have markings and appearances that mimic existing, proven and authentic products, but their quality, performance and safety are suspect or unknown.  Second, cloned tourniquet products have been seen in some countries.  While cloned products do not bear counterfeit markings, the fact that their physical appearance is similar to authentic products may lead users to mistakenly believe that their safety, quality and effectiveness will be equivalent.  This is not the case.  In considering comparative effectiveness, it is critical that surgical facilities and staff verify the origin and authenticity of tourniquet products.

 

Personalized Tourniquet Pressure Settings

It is well established by evidence in the clinical literature that higher tourniquet pressures are associated with higher probabilities of tourniquet-related injuries.  As a result, modern tourniquet systems aim to use the minimum pressure required to stop blood flow in a limb over the duration of a surgical procedure.  A new method based on Limb Occlusion Pressure (LOP) has been shown to allow individualized, optimal tourniquet pressure settings to be achieved.  LOP  can be defined as the minimum pressure required, at a specific time in a selected tourniquet cuff applied to an individual  patient’s limb at a desired location, to stop the flow of arterial blood into the limb distal to the cuff.   Some advanced surgical tourniquet systems include means to measure LOP automatically, although LOP can also be measured non-automatically by users.   Further information can be found at http://tourniquets.org/lop.php.

 

Personalized Tourniquet Cuffs

A recent introduction of personalized tourniquet cuffs has also resulted in safer and more effective tourniquet use.  Personalized cuffs are designed to better match patient limb size and shape and thus provide more efficient application of cuff pressure to the limb, allowing lower and safer tourniquet pressures to be used.  The improved fit is a result of the advent of new types of tourniquet cuff designs, in addition to the traditional tourniquet cuff design.  The traditional ‘straight’ tourniquet cuffs are best suited to cylindrical limb shapes.   New types of cuffs are ‘variable contour cuffs’ and allow the user to adapt the shape of the tourniquet cuff to any of a wide range of non-cylindrical (or tapered) limb shapes.  In addition the advent of new cuffs that allow better matching of cuff shapes to individual limb shapes, other advances in tourniquet cuff design have been made for pediatric and bariatric patient populations.  Tourniquet cuffs are now available that are matched specifically to pediatric and bariatric limb sizes and shapes, with comparative effectiveness established in published literature.

 

Reducing Soft Tissue Injuries with Matching Limb Protection Sleeves

High pressures, high pressure gradients and shear forces applied to skin and soft tissues underlying a tourniquet cuff can cause injuries to the skin and soft tissues.  To reduce the nature and extent of these injuries, studies have been published to determine the relative effectiveness of no protective material, underlying padding, underlying stockinette, and underlying limb protection sleeves that are matched to specific limb sizes and cuff sizes.  Study results present evidence that limb protection sleeves improve safety by protecting the skin underlying tourniquet cuffs during tourniquet use, and further provide evidence that greatest safety is achieved through the use of limb protection sleeves consisting of two-layer material specifically matched to the limb size and cuff size.  (See further information at www.tourniquets.org)

 

Reprocessing tourniquet cuffs

Outpatient surgical facilities are increasingly facing the question of whether, when, and how to reprocess tourniquet cuffs.  The answer requires consideration of patient safety, risk management, and cost.

For tourniquet cuffs designated as ‘reusable’ by manufacturers, the answers are straightforward because instructions on cleaning, inspecting and testing cuffs between uses are usually provided by the manufacturers.  Some facilities are reprocessing and reusing tourniquet cuffs designated as being ‘disposable’ or ‘single use’ by the manufacturers.  In such cases, no instructions on cleaning, inspecting and testing cuffs between uses are provided by the manufacturers.  For any facility considering the reprocessing and reuse of disposable or single-use cuffs, the following precautions should be taken.

  • A tourniquet cuff testing program should be established so that each cuff is thoroughly tested according to a written protocol after each use, and prior to the next use, with the results thoroughly documented.
  • A unique identifier should be used for each cuff so that the number of reuse cycles can be recorded.
  • Each cuff should be replaced after a maximum number of reuse cycles has been reached, to reduce the risk of cuff failure and patient injuries during use.
  • A tourniquet cuff testing protocol should be established in accordance with the recommendations of the original manufacturer of the cuff. At a minimum, the cuff testing protocol should include:

(a) a leak test, including inflating the cuff to a maximum pressure recommended by the manufacturer for a period of time, with the cuff wrapped around a test mandrel or laid flat,
(b) a fastener integrity test, to assure that the fasteners are not degraded to the point of being unsafe at the maximum pressure specified by the manufacturer,
(c) a physical inspection of the cuff to detect blockages of the pneumatic passageway in any portion of the inflatable bladder or tubing due to reprocessing damage or fluid entry,
(d) a visual inspection of the cuff to detect damage or deterioration, including: any warping of stiffener due to inappropriate reprocessing; discoloration or contamination of the cuff surface; damage or deterioration of the cuff connector or inflatable portion; and
(e) written documentation and evaluation of the test results before a decision is made regarding cuff reuse.

The complexity and cost of implementing an appropriate tourniquet cuff testing program may mean that the safe reprocessing of single-use tourniquet cuffs may not be cost-effective.

  

Emergency and military tourniquets

Some outstanding work by the US Army’s Institute for Surgical Research has led to the introduction and widespread use of tourniquets in combat settings.  It has been proven convincingly that many lives have been saved that would have been lost without the use of tourniquets.   As a result of these successes in combat settings, the same types of tourniquets are now being used increasingly by police, paramedics and other first responders in non-military settings with similar benefits.    Also, based on the proven safety and efficacy of pneumatic tourniquets in surgical settings over many years, new types of compact pneumatic tourniquets are being developed and used in emergency and military settings.  For example, a recent study of comparative effectiveness led to the introduction and use of pneumatic tourniquets by NATO forces. (See http://www.tourniquets.org for more information.)

Read Full Post | Make a Comment ( None so far )

Three Considerations When Buying Tourniquet Systems

Posted on October 25, 2013. Filed under: Instrument Selection | Tags: , , , , |

Jim McEwen PhD

Pneumatic tourniquets are used many thousands of times per day in orthopaedic and non-orthopaedic surgical procedures in the United States and elsewhere, facilitating operations by reliably establishing a bloodless surgical field with a high level of safety.

As with any type of medical device there is considerable variation between tourniquet systems currently available in the market in terms of cost, capabilities and safety features.  When considering a tourniquet system for purchase, buyers should take the following factors into consideration:

 

Safety

Within the last thirty years, there have been important improvements in the technology of tourniquet instruments and tourniquet cuffs.  Buyers should be aware of the types of safety features and alarms that may be included in the instruments and cuffs they are considering for purchase and carefully review the specifications of each.

Modern electronic tourniquet instruments include a pressure regulator that maintains cuff pressure at the level set by the user and an automatic timer to provide an accurate record of tourniquet inflation time and alarm if the inflation time exceeds a limit set by the user.  Tourniquet instruments also typically include audiovisual alarms to prompt the operator if hazardously high or low cuff pressures are present in the cuff.  Other safety related features and alarms that may be present in a tourniquet instrument include:

  • Self-test capabilities to provide automatic checks of system operation and calibration at each start-up of the instrument.
  • Self-monitoring capabilities to continuously monitor the operation of the instrument.
  • A backup battery to allow instruments to continue to operate normally during an unanticipated power interruption or during patient transport.
  • Alarms to detect potentially hazardous air leakage from pressurized tourniquet cuffs.
  • Alarms to detect occlusions of the tubing connecting the instrument to the cuff.
  • Alarms to detect failure of a cuff to depressurize when deflation is intended.
  • A cuff hazard interlock to prevent the instrument from being inadvertently powered off while a cuff is still inflated.
  • Interlocks to help prevent inadvertent cuff deflation during intravenous regional anesthesia (Bier Block) procedures.

Additional features that may be found in some of the most modern tourniquet systems include:

  • Automated estimation of Limb Occlusion Pressure, permitting individualized setting of safer tourniquet pressures.
  • Integrated cuff testing, the capability to test cuffs, tubing and connectors for leaks.
  • Interfaces to OR information systems to capture cuff pressures and inflation times.

In addition to the safety features listed above, the user interface of the tourniquet instrument may also include special safety features to help prevent inadvertent and unintended changes in cuff pressure by the user during a procedure.

The tourniquet cuffs that are applied to each patient should also be taken into consideration when selecting a tourniquet system.  Proper cuff selection and application is essential for safe tourniquet use.  Cuffs must provide adequate surgical exposure and properly fit the surgical limb.  Buyers should consider if cuffs are available for the instrument to correctly fit the patient population for which the system is intended.  Manufacturers offer cuffs in various sizes and shapes to fit a wide range of patient limbs, from the smallest pediatric patients to large bariatric patients.  Some manufacturers also provide matching limb protection sleeves to be used between the cuff and the limb to help prevent damage to the skin beneath the cuff.

 

Two tubes or one?

Tourniquet instruments and cuffs are commonly available in dual port (two tubes per cuff) or single port (one tube per cuff) configurations.  Dual port instruments and their corresponding cuffs have two separate tubing connections between the instrument and each attached cuff.  One connection regulates the air pressure in the cuff and the other monitors the pressure in the cuff.  Generally speaking, dual port instruments and cuffs provide the most accurate and reliable indication of pressure within the cuff.  This is especially important when the tourniquet cuff is used to measure Limb Occlusion Pressure as described above.

Single port instruments and their corresponding cuffs have a single tube connecting the instrument to the cuff.  The connection is used for both pressure regulation and monitoring.  Single port instruments and cuffs usually cost less than dual port instruments.

Adaptors should not be used to connect single port cuffs to dual port instruments or to connect dual port cuffs to single port instruments.

 

One cuff or two?

Tourniquet instruments are available in single and dual channel configurations.  Dual channel instruments can control the pressure in two cuffs simultaneously, facilitating bilateral and IVRA (Bier Block) procedures.  Each cuff channel should have independent indicators for cuff pressure and inflation time and independent, well identified controls for cuff inflation and deflation.  The controls, indicators and cuff connecting tubing should be color coded to help identify the cuff connected to each channel.  For additional safety, interlocks are included in some systems to help prevent inadvertent cuff deflation during IVRA procedures.

Single channel instruments control the pressure in only one cuff.  They are generally smaller and less costly than dual channel instruments.  Two single channel instruments may be used together for bilateral and IVRA procedures.  Under no circumstances should adaptors be used to connect two or more cuffs to a tourniquet instrument intended for single channel operation.

When selecting a tourniquet instrument and cuff system buyers should always consider their clinical needs and carefully review the safety features and performance specifications of the systems under consideration.

Read Full Post | Make a Comment ( None so far )

Evidence of safety: matching limb protection sleeves with tourniquet cuffs

Posted on March 4, 2013. Filed under: Matching Limb Protection Sleeves, Tourniquet Safety | Tags: , , , , , |

What evidence exists to show the extent to which the use of limb protection sleeves underlying tourniquet cuffs improves patient safety?  It is known that high pressures, high pressure gradients and shear forces applied to skin and soft tissues underlying a tourniquet cuff can cause injuries to the skin and soft tissues including petechiae, blistering, bruising, and pinching.  To reduce the nature and extent of such injuries, studies have been published to determine the relative effectiveness of underlying padding, underlying stockinette, and underlying limb protection sleeves that are matched to specific limb sizes and cuff sizes.

Olivecrona et al. [1] compared the use of elastic stockinette, cast padding, and no protective material underneath a pneumatic tourniquet in 92 patients who underwent a primary total knee arthroplasty and confirmed that an elastic stockinette provided the most effective skin protection during tourniquet use, of the three that were studied.  McEwen et al. [2] compared five different tourniquet cuff sleeve and padding configurations in healthy adult volunteers, and Tredwell et al. [3] tested four different tourniquet cuff sleeve and padding configurations on healthy child volunteers.  Both studies found that stretched sleeves made of two-layer tubular elastic material and matched to specific tourniquet cuff sizes, and thus to limb sizes, produced significantly fewer, less severe pinches and wrinkles in the skin surface than other types of underlying limb protection that were tested, including cast padding, single-layer stockinette, and stockinette not matched to the cuff and limb size.

These studies present evidence that limb protection sleeves improve safety by protecting the skin underlying tourniquet cuffs during tourniquet use, and further provide evidence that greatest safety is achieved through the use of matching limb protection sleeves consisting of two-layer tubular stockinette specifically matched to the limb size and cuff size.  Detailed information about matching limb protection sleeve products is available elsewhere [4].

References

[1]      J. A. McEwen and K. Inkpen, “Tourniquet Safety: Preventing Skin Injuries,” The Surgical Technologist, no. August 2002, pp. 7–15, 2002.

[2]      S. J. Tredwell, M. Wilmink, K. Inkpen, and J. a McEwen, “Pediatric tourniquets: analysis of cuff and limb interface, current practice, and guidelines for use.,” Journal of Pediatric Orthopedics, vol. 21, no. 5, pp. 671–6, 2001.

[1]      C. Olivecrona, J. Tidermark, P. Hamberg, S. Ponzer, and C. Cederfjäll, “Skin protection underneath the pneumatic tourniquet during total knee arthroplasty: a randomized controlled trial of 92 patients.,” Acta Orthopaedica, vol. 77, no. 3, pp. 519–23, Jun. 2006.

[4]        http://www.delfimedical.com/pts-ii/limb-sleeves

Read Full Post | Make a Comment ( None so far )